Test Code HCG S Qual   hCG Qualitative Serum Pregnancy

Performing Laboratory

Memorial Hospital Laboratory- Automated Procedures  304-388-5953       
General Hospital Laboratory 304-388-6244
Women and Children's Hospital Laboratory 304-388-2385
Teays Valley Hospital Laboratory 304-757-1770
Plateau Medical Center Laboratory 304-469-8621
Greenbrier Valley Laboratory 304-647-6060

Specimen Requirements

Serum-Plain Red Top Tube or SST

Draw 4 ml (minimum 1 microtainer) 

Allow sample to clot naturally before centrifugation

Centrifuge and remove serum within 8 hours of collection    

Specimen Rejection

  • Improperly labeled, collected, stored, and transported
  • Quantity not sufficient
  • Specimen other than serum

Storage and Stability

Refrigerate 2-8C              48 hours
Frozen < -20C Prolonged Storage             

Reference Values

Negative 

Methodology

Rapid chromatographic immunoassay

Days Test Set Up and Turnaround

Monday through Sunday, Continuously

AM Collection Expected by 9AM same day
Routine Expected within 24 hrs
Stat Expected within 1 hr
Timed Expected results are test dependant

CPT Coding

84703

Additional Information

Memorial, General, WCH: The SP Brand Rapid test detects hCG at concentrations of 10 mIU/mL or greater.  The test has a sensitivity of 10 mIU/mL, and is capable of detecting pregnancy as early as 1 day after the first missed menses.  False negatives results may occur when the levels of hCG are below the sensitivity level of the test.  When pregnancy is still suspected, a first morning urine or serum sample should be collected 48 hours later.

Teays Valley Hospital:  Uses the Beckman Coulter ICON 25 hCG test which is a rapid test that qualitatively detects the presence of hCG in urine samples at the sensitivity of >/=25 mIU/mL