Test Code HIT (PF4) Ab Heparin-PF4 Antibody (HIT)
Performing Laboratory
Memorial Hospital Laboratory- Automated Procedures | 304-388-5953 |
Specimen Requirements
Plasma- Light Blue Top, Sodium Citrate
Draw 1 full, tube by “clean” venipuncture (no excessive trauma).
Immediately after draw, invert tubes several times to mix blood with additive.
Specimen Rejection
- Improperly labeled, collected, stored, and transported
- Quantity not sufficient
- Hemolyzed
- Marked lipemia
- Clotted specimens
Storage and Stability
Room temperature | 4 hours |
Frozen < -20oC | 2 weeks |
Reference Values
Negative
Interpretation
Detects unbound antibody only. If antibody is bound due to adminsitration of heparin, it will not be detected. Heparin induced thrombocytopenia is not diagnosed by this test. Antibodies may be present without the clincial syndrome and the clinical syndrome may be present without detectable antibodies.
Methodology
Latex Immumoassay
Days Test Set Up and Turnaround
Monday through Sunday, Dayshift Only, 7:00 AM-2:30PM
Turnaround Time-24 Hours
CPT Coding
86022