Test Code PLT Platelet Count
Performing Laboratory
| Memorial Hospital Laboratory- Automated Procedures | 304-388-5953 |
| General Hospital Laboratory | 304-388-6244 |
| Women and Children's Hospital Laboratory | 304-388-2385 |
| Teays Valley Hospital Laboratory | 304-757-1770 |
| Cancer Center Laboratory | 304-388-831 |
| Plateau Medical Center Laboratory | 304-469-8621 |
| Greenbrier Valley Laboratory | 304-647-6060 |
Specimen Requirements
Whole Blood-Lavender-top (EDTA) tube or microtainer
3 mL of whole blood (minimum 1 full microtainer)
Specimen Rejection
- Improperly labeled, collected, stored, and transported
- Quantity not sufficient
- Clotted specimen
- Grossly hemolyzed
- Contaminated
Storage and Stability
| Room Temperature 18-24o C | 24 hours |
| Refrigerate 2-8o C | 48 hours |
| Frozen | Unacceptable |
Specimens analyzed on back up instruments at Cancer Center, Teays Valley or Women's and Childrens cannot be analzed if refrigerated longer than 8 hours.
Reference Values
| Male | Female | Male | Female | M/F | M/F | M/F | |
| >18yo | >18yo | 10-18yo | 10-18yo | 2-10yo | 1mon-2yr | 0-1 month | |
| x103/mcL | 140-450 | 140-450 | 125-365 | 126-404 | 163-461 | 188-506 | 137-475 |
Critical Value: < 30 x103/mcL or > 1000 x103/mcL
Methodology
Coulter principle or hemacytometer
Days Test Set Up and Turnaround
Monday through Sunday, Continuously
| AM Collection | Expected by 9AM same day |
| Routine | Expected within 24 hrs |
| Stat | Expected within 1 hr |
| Timed | Expected results are test dependant |
CPT Coding
85049
Additional Information
A blue top (Sodium Cirate tube) may be collected due to EDTA causing platelets to clump.